RecallLookupBack to the product checker
Serious riskSerious risk

EU Safety Gate · Report-2023-3

Lifa air · Particle filter mask

The particle/filter retention of the material is insufficient (measured values as low as 46%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. The product does not comply with the Personal Protective Equipment Regulation nor with the European standard EN 149.

Published
20 January 2023
Notifying authority
Finland
Catalog checked
30 July 2026
What should you do?

Check the authority measure and original alert.

Type of economic operator to whom the measure(s) were ordered: Manufacturer. Category of measure(s): Recall of the product from end users. Date of entry into force: 16/11/2022 · Type of economic operator to whom the measure(s) were ordered: Manufacturer. Category of measure(s): Withdrawal of the product from the market. Date of entry into force: 16/11/2022 · Type of economic operator to whom the measure(s) were ordered: Manufacturer. Category of measure(s): Stop of sales. Date of entry into force: 16/11/2022

Official source alert
Safety Gate reference
A12/00139/23
Model
FFP2 NR LMp300x
Batch
41/21
Check-digit validated barcodes
6430077231412
Category
Protective equipment
Risk type
Health risk / other
Country of origin
Finland

Respiratory half mask, available in different colours, with CE marking.