EU Safety Gate · Report-2021-29
Univent medical · atemious pro
The total filtration capacity of the mask and the particle/filter retention of the material are insufficient (measured mean values as low as 78%, and measured filter penetration values as low as 89%, respectively). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. The product does not comply with the Personal Protective Equipment (PPE) Regulation and with the relevant European standard EN 149.
- Published
- 23 July 2021
- Notifying authority
- Germany
- Catalog checked
- 30 July 2026
Check the authority measure and original alert.
Type of economic operator to whom the measure(s) were ordered: Other. Category of measure(s): Recall of the product from end users. Date of entry into force: Unknown. Type of economic operator taking notified measure(s): Manufacturer. Category of measure(s): Stop of sales. Date of entry into force: Unknown
Official source alertProduct identification
- Safety Gate reference
- A12/01064/21
- Model
- 2001
- Check-digit validated barcodes
- 4260688380047
Classification
- Category
- Protective equipment
- Risk type
- Health risk / other
- Country of origin
- Germany
Product description
White respiratory face mask with ear loops. The product was sold online, in particular via Amazon.