EU Safety Gate · Report-2021-46
PRINTEXMedical · priuMask
The particle/filter retention of the material is insufficient (measured values: as low as 84%). Furthermore, the mask may lack a nose clip, therefore it cannot be correctly adapted to the user's face. Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. The product does not comply with the Personal Protective Equipment (PPE) Regulation and with the relevant European standard EN 149.
- Published
- 19 November 2021
- Notifying authority
- Germany
- Catalog checked
- 30 July 2026
Check the authority measure and original alert.
Type of economic operator to whom the measure(s) were ordered: Other. Category of measure(s): Ban on the marketing of the product and any accompanying measures. Date of entry into force: 26/08/2021 · Type of economic operator to whom the measure(s) were ordered: Other. Category of measure(s): DIN EN 149: Respiratory protective devices – Filtering half-masks to protect against part. Date of entry into force: 26/08/2021 · Type of economic operator taking notified measure(s): Manufacturer. Category of measure(s): Enclose required documents. Date of entry into force: 26/08/2021 · Type of economic operator taking notified measure(s): Manufacturer. Category of measure(s): Destruction of the product. Date of entry into force: 26/08/2021 · Type of economic operator taking notified measure(s): Manufacturer. Category of measure(s): Withdrawal of the product from the market. Date of entry into force: 26/08/2021
Official source alertProduct identification
- Safety Gate reference
- A12/01580/21
- Model
- P21H1
- Batch
- 21.12.2020
Classification
- Category
- Protective equipment
- Risk type
- Health risk / other
- Country of origin
- Not published
Product description
5-pack of respiratory face masks branded as FFP2, in a transparent foil packaging.