RecallLookupBack to the product checker
Serious riskSerious risk

EU Safety Gate · Report-2020-52

Welooo · KN95 Filtering half mask

The particle/filter retention of the material is insufficient (measured value 44,1%). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. The product does not comply with the Personal Protective Equipment Regulation and with the relevant European standard EN 149.

Published
25 December 2020
Notifying authority
Germany
Catalog checked
30 July 2026
What should you do?

Check the authority measure and original alert.

Type of economic operator taking notified measure(s): Importer. Category of measure(s): Destruction of the product. Date of entry into force: 07/12/2020 · Type of economic operator taking notified measure(s): Retailer. Category of measure(s): Recall of the product from end users. Date of entry into force: 25/11/2020

Official source alert
Safety Gate reference
A12/01886/20
Model
KN95 / FFP2
Batch
20200417
Check-digit validated barcodes
6973109940017
Category
Protective equipment
Risk type
Health risk / other
Country of origin
People's Republic of China

Single use particle half-face filter mask for protection against particles and aerosols, labelled KN95 and FFP2. The product is embossed with "KN95" on the side.