RecallLookupBack to the product checker
Serious riskSerious risk

EU Safety Gate · Report-2019-031

AEDEN · SEMICONDUCTOR LASER TREATMENT INSTRUMENT

The laser beam is too powerful. Direct viewing of the laser beam could cause damage to sight. The product does not comply with the relevant European Standard EN 60825-1.

Published
02 August 2019
Notifying authority
Slovakia
Catalog checked
30 July 2026
What should you do?

Check the authority measure and original alert.

Type of economic operator to whom the measure(s) were ordered: Retailer. Category of measure(s): Withdrawal of the product from the market. Date of entry into force: 21/04/2019 · Type of economic operator to whom the measure(s) were ordered: Retailer. Category of measure(s): Warning consumers of the risks. Date of entry into force: 21/04/2019 · Type of economic operator to whom the measure(s) were ordered: Retailer. Category of measure(s): Recall of the product from end users. Date of entry into force: 21/04/2019

Official source alert
Safety Gate reference
A12/1144/19
Model
GD07-W
Category
Other - -
Risk type
Damage to sight
Country of origin
People's Republic of China

Semiconductor laser phototherapy device. The product includes a host machine and а nasal treatment probe with a laser. Packed in a box.