RecallLookupBack to the product checker
Official safety alertOther risk levels

EU Safety Gate · Report-2025-19

TRANS MEDIKAL - GETINGE · Steam generator

Due to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns. The product does not comply with the requirements of the Pressure Equipment Directive.

Published
09 May 2025
Notifying authority
France
Catalog checked
30 July 2026
What should you do?

Check the authority measure and original alert.

Type of economic operator to whom the measure(s) were ordered: Distributor. Category of measure(s): Recall of the product from end users. Date of entry into force: 02/03/2025 · Type of economic operator taking notified measure(s): Distributor. Category of measure(s): Making the marketing of the product subject to prior conditions. Date of entry into force: 02/03/2025

Official source alert
Safety Gate reference
OR/00067/25
Model
20 kW ; 22,5 kW ; 30 kW ; 37,5 kW ; 40 kW ; 45 kW ; 50 kW ; 60 kW
Category
Pressure equipment/vessels
Risk type
Burns
Country of origin
Türkiye

Steam generator integrated into steriliser.