EU Safety Gate · Report-2025-19
TRANS MEDIKAL - GETINGE · Steam generator
Due to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns. The product does not comply with the requirements of the Pressure Equipment Directive.
- Published
- 09 May 2025
- Notifying authority
- France
- Catalog checked
- 30 July 2026
Check the authority measure and original alert.
Type of economic operator to whom the measure(s) were ordered: Distributor. Category of measure(s): Recall of the product from end users. Date of entry into force: 02/03/2025 · Type of economic operator taking notified measure(s): Distributor. Category of measure(s): Making the marketing of the product subject to prior conditions. Date of entry into force: 02/03/2025
Official source alertProduct identification
- Safety Gate reference
- OR/00067/25
- Model
- 20 kW ; 22,5 kW ; 30 kW ; 37,5 kW ; 40 kW ; 45 kW ; 50 kW ; 60 kW
Classification
- Category
- Pressure equipment/vessels
- Risk type
- Burns
- Country of origin
- Türkiye
Product description
Steam generator integrated into steriliser.